Quality Control
Quality control is designed to detect, reduce, and correct deficiencies in a laboratory's internal analytical process prior to the release of patient results.Quality command samples are special specimens inserted into the testing process and treated as if they were patient samples by being exposed to the same operating weather condition. The purpose of including quality control samples in analytical runs is to evaluate the reliability of a method by assaying a stable material that resembles patient samples.Quality control is a measure out of precision or how well the measurement system reproduces the aforementioned result over fourth dimension and under varying operating conditions.Pathologists need to be involved in development of quality control protocols, the selection of quality control materials, long term review of quality command data, and decisions about repeating patient samples later on large runs are rejected.These quality command activities play an important office in assuring the quality of laboratory tests.
Quality command cloth is usually run at the first of each shift, after an instrument is serviced, when reagent lots are changed, subsequently scale, and when patient results seem inappropriate.A quality command scheme must be developed that minimizes reporting of erroneous results, but does not result in excessive repitition of analytical runs. The manufacturer should recommend in their production labeling the period of time within which the accuracy and precision of the instruments and reagents are expected to exist stable.Each laboratory should employ this information to determine their belittling run length, taking into consideration sample stability, reporting intervals of patient results, cost of reanalysis, work flow patterns, and operator characteristics.The user'southward defined run length should not exceed 24 hours or the manufacturer's recommended run length.Quality control samples must be analyzed at least in one case during each belittling run.Manufacturers should recommend the nature of quality command specimens and their placement inside the run.Random placement of quality command samples yields a more valid approximate of belittling imprecision of patient data than fixed placement and is preferable.
Quality control materials should have the following characteristics.They should have the same matrix as patient specimens, including viscosity, turbidity, composition, and color.For example, a method that assays serum samples should be controlled with human serum based controls. Quality command textile should exist uncomplicated to use because complicated reconstitution procedures increase the take a chance of error.Liquid controls are more user-friendly than lyophilized controls because they exercise non take to exist reconstituted.Controls should have minimal vial to vial variability, considering variability could be misinterpreted as systematic fault in the method or instrument.Quality control materials should be stable for long periods of fourth dimension.Controls with short shelf lives necessitate frequent reordering and verification against the outgoing material, creating more than unnecessary work.Quality control material should be available in large enough quantities to concluding at least ane year. Purchasing a large batch decreases the number of times that control ranges have to be established.
Controls should have target values that are close to medical decision points.Quantitative tests should include a minimum of one control with a target value in the healthy person reference interval and a 2d control with a target value that would be seen in a sick patient.Examples include sodium controls of 140 and 115 mEq/50 and glucose controls of 75 and 225 mg/dL. If 3 control levels are run, an abnormally low patient range should be included.Quality command levels for therapeutic drug monitoring should mirror therapeutic, toxic, and trough values.If a test is qualitative, giving either negative or positive results, a negative command and a weak positive command with a concentration at the lowest detectable level are recommended.Semi-quantitative tests should have controls at each graded level - trace, ane+, 2+, etc.
Both assayed and unassayed control material are available.Assayed controls are measured past a reference method and sold with published target values.They are more expensive than unassayed controls and are non toll effective for routine quality control in a hospital or reference laboratory.Assayed controls are recommended for physician office laboratories.Unassayed controls must be analyzed past the laboratory to make up one's mind the target value and acceptable range.Comparison studies demand to be run between the current and new unassayed control materials.If the new control fabric is from the same manufacturer, but v samples of the new control cloth need to be run to constitute a hateful.If the mean is close to the hateful of the approachable quality command fabric, the new control textile tin can be accepted.No information points should be excluded unless they are known to be result of operational errors.The standard deviation of the outgoing controls is adopted for employ until plenty data points are collected for calculation..
Interpretation of quality control data involves both graphical and statistical methods. Quality control data is most easily visualized using a Levey-Jennings command nautical chart. The dates of analyses are plotted along the 10-axis and command values are plotted on the Y-axis.The mean and one, two,and iii standard deviation limits are as well marked on the Y-axis.Inspecting the pattern of plotted points provides a simple way to find increased random error and shifts or trends in scale.With a correctly operating system, repeat testing of the same control sample should produce a Gaussian distribution.That is, approximately 66% of values should fall between the +/- 1 s ranges and be evenly distributed on either side of mean.Ninety five percent of values should lie between the +/- 2 s ranges and 99% between the +/- iii due south limits.This means that 1 data point in 20 should autumn between either of the 2 south and 3 due south limits and one data point in 100 volition autumn outside the 3 s limits in a correctly operating system.In full general, the +/- 2 s limits are considered to be alert limits.Values falling between 2 s and 3 s indicates the analysis should be repeated.The +/-3 s limits are rejection limits.When a value falls outside of these limits the assay should stop,patient results held, and the test system investigated.
Normal Distribution of Control Values
+3s | ||||||||||||||||||||
+2s | x | |||||||||||||||||||
+1s | 10 | x | x | x | x | |||||||||||||||
u | ten | x | ten | x | x | ten | ten | |||||||||||||
-1s | ten | x | x | x | ten | |||||||||||||||
-2s | x | ten | ||||||||||||||||||
-3s | ||||||||||||||||||||
Day | 01 | 02 | 03 | 04 | 05 | 06 | 07 | 08 | 09 | 10 | xi | 12 | 13 | xiv | 15 | 16 | 17 | eighteen | nineteen | xx |
Reviewing the pattern of points plotted over time is useful in spotting shifts and trends in method calibration.A shift is a sudden change of values from one level of the control chart to another.A mutual cause of a shift is failure to recalibrate when changing lot numbers of reagents during an belittling run.A tendency is a continuous movement of values in one direction over vi or more analytical runs.Trends tin start on one side of the mean and motility across information technology or tin occur entirely on i side of the mean.Trends tin can be caused by deterioration of reagents, tubing, or light sources.Shifts and trends can occur without loss of precision and tin can occur together or independently.The occurrence of shifts and trends on the Levey-Jennings control chart is the effect of either proportional or constant mistake.
A Shift in Control Values
+3s | x | x | ||||||||||||||||||
+2s | x | 10 | x | x | x | x | x | |||||||||||||
+1s | x | 10 | x | x | ||||||||||||||||
u | 10 | ten | x | |||||||||||||||||
-1s | x | ten | ||||||||||||||||||
-2s | ten | x | ||||||||||||||||||
-3s | ||||||||||||||||||||
Day | 01 | 02 | 03 | 04 | 05 | 06 | 07 | 08 | 09 | 10 | 11 | 12 | 13 | xiv | xv | sixteen | 17 | 18 | 19 | 20 |
A Tendency in Control Values
+3s | ||||||||||||||||||||
+2s | ||||||||||||||||||||
+1s | x | x | ||||||||||||||||||
u | x | 10 | ||||||||||||||||||
-1s | x | x | x | x | ||||||||||||||||
-2s | ten | ten | x | x | x | |||||||||||||||
-3s | x | x | ten | 10 | x | x | ten | |||||||||||||
Day | 01 | 02 | 03 | 04 | 05 | 06 | 07 | 08 | 09 | ten | xi | 12 | 13 | 14 | xv | sixteen | 17 | xviii | 19 | 20 |
Levy-Jennings charts tin also demonstrate loss of precision by an increment in the dispersion of points on the command chart.Values can remain within the +/-2 southward and
3 s limits, just exist unevenly distributed outside of the +/-1 southward limits.Random error is present if more than one in twenty values fall across the +/-2 southward limits.
Increased Dispersion of Command Values
+3s | ten | x | x | |||||||||||||||||
+2s | 10 | 10 | ||||||||||||||||||
+1s | x | x | x | |||||||||||||||||
u | x | 10 | x | 10 | x | |||||||||||||||
-1s | x | ten | ||||||||||||||||||
-2s | x | x | ||||||||||||||||||
-3s | x | x | x | |||||||||||||||||
Twenty-four hour period | 01 | 02 | 03 | 04 | 05 | 06 | 07 | 08 | 09 | 10 | 11 | 12 | xiii | xiv | 15 | 16 | 17 | 18 | 19 | 20 |
Past running and evaluating the results of 2 controls together, trends and shifts can be detected much earlier.Westgard and associates have formulated a serial of multi-rules to evaluate paired control runs.Westgard rules are:
- 12s rule - one command value is exterior a 2 standard deviation limit from the hateful.Failure of this dominion does not always indicate an analytical mistake has occurred.If the control falls outside of the 2s limit considering of a normal Gaussian distribution, the adjacent control consequence should pass this dominion.
- 13s rule - ane control value falls outside a three standard deviation limit from the mean.This may exist the result of random mistake and should be investigated.
- 22s rule - two consecutive values have fallen outside of the same 2s limit.This rule tin utilize to a single level during ii consecutive runs or both levels of control during the same run.Violation suggests systematic error.
- R4s dominion - the range between two consecutive values of the same control level is greater than or equal to 4 standard deviations.This rule besides applies between control levels.I control is beyond the +2s limit and the other is beyond the -2s limit.Violation suggests random fault.
- 41s rule - four consecutive values are accept fallen on the same side of the same 1s range.This dominion can involve one or both control levels.Violation suggests systematic error.
- 10ten rule - ten sequent values fall on the same side of the hateful. This rule can apply within i control level or between control levels.Failure of this rule indicates a shift and the presence of systematic error.
Westgard rules are programmed in to automatic analyzers to decide when an analytical run should be rejected.These rules demand to exist practical carefully then that true errors are detected while imitation rejections are minimized.The rules applied to high volume chemistry and hematology instruments should produce low fake rejection rates.A single rule with a big limit, such as the 13s rule, is recommended.Batch analyzers and transmission tests require command samples in each batch.If the test is error prone, the quality control protocol needs to have a loftier error detection rate.Mistake detection is improved by increasing the number of controls per run, narrowing control acceptance limits, and using Westgard rules with tighter limits.
Analytes with big biological (intra-private) variation do not crave as much analytical accuracy as analytes with pocket-size biological variations.One recommendation is that full analytical variation should be less than half the biological variation (see Appendix A for a consummate listing of biological variation).For case, the biological variation of fasting triglycerides is ~20%; therefore, analytical variation tin can be as high every bit ten% without significantly affecting medical determination making.Examples of recommended goals for imprecision (%CV) for some commonly ordered chemistry tests at their medical decision points are listed beneath.
Examples of Analytical Goals for Method Precision (CV %)
Analyte | Method CV % | Analyte | Method CV % |
Albumin | 1.4 | IgA | two.two |
ALT | 13.6 | IgG | 1.9 |
ALP | 3.4 | IgM | two.3 |
AST | 7.two | Atomic number 26 | 15.9 |
Amylase | 3.7 | Phosphate | 4.0 |
Bicarbonate | 2.3 | Potassium | 2.four |
Bilirubin | 11.3 | Proteins | 1.four |
Calcium | 0.9 | Sodium | 0.3 |
Chloride | 0.7 | T4 | 3.iii |
Cholesterol | two.7 | TSH | 8.1 |
Creatinine | ii.2 | Urate | 4.ii |
Glucose | 2.2 | Urea | vi.3 |
Some of the most common problems causing quality control samples to shift are summarized in the following table.
Shift within range | Shift out of range | Trending | Excessive besprinkle |
Improper mixing of controls | Any of column 1 | Change in musical instrument reaction temperature | Improper mixing in instrument |
Controls left at room temperature also long | Improper reconstitution of control | Musical instrument sampling problem | Contamination during testing |
Vial to vial variation | Concentration of control in fault | Instrument reagent commitment problem | |
Alter in reagent lot number (specially with enzymes) | Contaminated reagent | ||
Control deteriorated | |||
Musical instrument malfunction |
When these bug are identified the following cosmetic actions can be taken.
- Check expiration date of the control.
- Check expiration date of the reagent.
- If a new command was used, brand sure information technology was reconstituted properly.
- Retest the control.If the new value is within adequate limits, record both values and proceed with patient testing.The trouble with the kickoff value was probably random fault, which is expected in one of every 20 values.
- If the repeat value is notwithstanding out of range, run a new vial of control.If the new control value is within acceptable limits, record the values and continue with patient testing.The problem with the starting time set of controls was probably specimen deterioration.
- If the new control value is out of control, troubleshoot the musical instrument (check sampling, reagent delivery, mixing, lamp integrity, and reaction temperature).
- Recalibrate the method, specially if two or more controls accept shifted.
- If controls shift after a new reagent lot number has been introduced, rerun some normal and aberrant patient samples.If patient correlations are practiced, control shifts are probably acceptable.If they are poor, reagent may exist bad.
- Attempt a new lot number of reagent.If the problem is corrected, cheque with the manufacturer to find out if anyone else has reported problems.
It is of import to record every quality command value, including those that are out of control.The object of quality control is not to produce beautiful command charts by trying to keep all results within +/- 2 SD.5 percent of control values are expected to be out of range.If quality command samples are routinely rerun until they fall within the electric current control limits and the outliers are not recorded, the acceptable range will become smaller each time they are recalculated.Eventually they volition approach zero standard deviation and go useless and unattainable.
When an analytical run is rejected because quality control is out of acceptable limits, it is often necessary to determine if patient results reported betwixt the terminal adequate run and the rejected run demand to be repeated.This decision should be based on the nature and size of the error.A 5% bias has no clinical significance for most patients, simply a 25% bias is unacceptable for nearly all patients.Biases in betwixt these extremes demand to be examined on a case by case ground.If analytical errors have clinical significance, and then some of the patient specimens should be re-tested, starting with the samples analyzed merely earlier the rejected control samples.
For example, suppose all glucose results in a run had a 10% negative bias.A patient with a true claret glucose of 82 mg/dL would take been reported as 74 mg/dL.Both values are normal and the results do not demand to be corrected.However, an fault of this magnitude would be significant for patient samples near medical decision points such as values below 60 mg/dL, fasting glucose values nearly 140 mg/dL, and glucose tolerance tests nearly 180 mg/dL.An acceptable strategy would exist to repeat patient samples that were less than threescore or greater than 140 mg/dL, starting with samples analyzed just earlier the quality command failure.Repeat testing should be washed in reverse chronological club until the new glucose results closely friction match the original results.